分析型高效液相色谱仪 Analytical HPLC

  • 41/人
    使用者
  • 487/次
    机时次数
  • 1806.62/小时
    总时长
  • 0/次
    送样次数
  • 91/人
    收藏者

收费标准

机时
0元/小时

设备型号

Agilent 1260 Infinity II

当前状态

管理员

文美玲 Wendy Meiling WEN,郑婷 020-88338595

放置地点

E4-126/133走廊-132
  • 仪器信息
  • 预约资源
  • 附件下载
  • 公告
  • 同类仪器

名称

分析型高效液相色谱仪 Analytical HPLC

资产编号

BC-PM-0006

型号

Agilent 1260 Infinity II

规格

产地

厂家

Agilent Technologies

所属品牌

Agilent

出产日期

购买日期

所属单位

生物科学中央实验室 Biosciences Central Research Facility (GZ)

使用性质

科研

所属分类

BioCRF-蛋白质&代谢组学平台 Proteomics & Metabolomics Core

资产负责人

--

联系电话

020-88338595

联系邮箱

meilingwen@hkust-gz.edu.cn

放置地点

E4-126/133走廊-132
  • 主要规格&技术指标
  • 主要功能及特色
  • 设备使用相关说明
  • 备注
主要规格&技术指标
G7129A 1260 Infinity II Vialsampler
Weight: 19 kg (41.9 lbs) w/o sample thermostat
Dimensions (height × width ×depth): 320 x 396 x 468 mm (12.8 x 15.6 x 18.4 inches)
Line voltage:100 – 240 V~, ± 10 %
Line frequency: 50 or 60 Hz, ± 5 %
Power consumption: 350 VA, 350 W, 1195 BTU/h
Ambient operating temperature: 4 - 40 °C (39 - 104 °F); without sample
Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
Humidity: < 95 % r.h. at 40 °C (104 °F)1
Operating altitude: Up to 3000 m (9842 ft)
Safety standards: IEC, EN,CSA, UL: Overvoltage category II, Pollution degree 2
ISM Classification: ISM Group 1 Class B According to CISPR 11
Permitted solvents: Boiling point ≥56 °C (133 °F). Auto-ignition temperature ≥200 °C
(392 °F).

G7111B Agilent 1260 Infinity II Quaternary Pump
Weight: 14.5 kg (32 lbs)
Dimensions (height × width × depth):180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches)
Line voltage: 100 – 240 V~, ± 10 %
Line frequency: 50 or 60 Hz, ± 5 %
Power consumption: 80 VA, 65 W
Ambient operating temperature: 4–55 °C (39–131 °F)
Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
Humidity: < 95 % r.h. at 40 °C (104 °F)
Operating altitude: Up to 3000 m (9842 ft)
Safety standards:IEC, EN, CSA, UL: Overvoltage category II, Pollution degree 2
ISM Classification: ISM Group 1 Class B According to CISPR 11

G7114A Agilent 1260 Infinity II Variable Wavelength Detector
Weight: 11 kg (24.3 lbs)
Dimensions (height × width × depth): 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches)
Line voltage: 100 – 240 V~, ± 10 %
Line frequency: 50 or 60 Hz, ± 5 %
Power consumption: 80 VA, 70 W
Ambient operating temperature: 4 - 55 °C (39 - 131 °F)
Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
Humidity: < 95 % r.h. at 40 °C (104 °F)
Operating altitude: Up to 3000 m (9842 ft)
Safety standards:IEC, EN, CSA, UL: Overvoltage category II, Pollution degree 2
ISM Classification: ISM Group 1 Class B

G7116A Agilent 1260 Infinity II Multicolumn Thermostat
Weight: 12.5 kg (27.6 lbs)
Dimensions (height × width × depth): 160 x 435 x 436 mm (6.3 x 17.1 x 17.2 inches), Width with column identification kit:460 mm
Line voltage: 100 – 240 V~, ± 10 %
Line frequency: 50 or 60 Hz, ± 5 %
Power consumption: 150 VA, 150 W
Ambient operating temperature: 4–55 °C (39–131 °F)
Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
Humidity: < 95 % r.h. at 40 °C (104 °F)
Operating altitude Up to 3000 m (9842 ft)
Safety standards:IEC, EN, CSA, UL: Overvoltage category II, Pollution degree 2
ISM Classification: ISM Group 1 Class B

G7121A Agilent 1260 Infinity II Fluorescence Detector
Weight: 11.9 kg (26.2 lbs)
Dimensions (height × width × depth): 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches)
Line voltage: 100 – 240 V~, ± 10 %
Line frequency: 50 or 60 Hz, ± 5 %
Power consumption: 70 VA / 60 W
Ambient operating temperature: 4-40 °C (39-104 °F)
Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
Humidity: < 95 % r.h. at 40 °C (104 °F)
Operating altitude: Up to 3000 m (9842 ft)
Non-operating altitude: Up to 4600 m (15092 ft)
Safety standards: IEC, EN, CSA, UL: Installation category II, Pollution degree 2
ISM Classification: ISM Group 1 Class B
主要功能及特色
G7129A 1260 Infinity II Vialsampler
• Capacity of up to 132 vials (2 mL) or up to 36 vials (6 mL)
• Reliable injections from 0.1 to 100 µL at up to 600 bar
• Injection volume extension up to 1,800 µL for microbore to semipreparative applications
• Needle flush port enables rinsing the outside surfaces of the needle for lowest carryover
• Integrated sample thermostat, provides cooling and heating from 4 to 40 °C
• Low internal volume for minimum contribution to system's total internal volume, which can be even further reduced using bypass mode
• Overlapped sample runs and injection cycles for increased productivity
• Customizable injection program for advanced injection sequences or sample preparation prior to injection

G7111B Agilent 1260 Infinity II Quaternary Pump
• Higher resolution, faster separations - the maximum pressure of 600 bar allows you to use smaller particle size columns, longer columns and/or higher viscous solvents as an alternative to acetonitrile.
• Analytical and semi-preparative purifications - the wide flow range (up to 10 mL/min) perfectly supports standard and semi-preparative applications. The perfect choice for 4.6 mm and 3.0 mm id columns.
• Quicker experiments - the pump offers convenient access to up to four solvents for isocratic or gradient analysis which can really speed up method development, preparation of mobile phases and system flushing.
• Easy, fast and secure transfer of methods from HPLC to sub-2-micron columns.
• Integrated degasser
• Longevity and low operating costs - robust materials (stainless steel, titanium, gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most demanding applications and keep your pump functioning for years. Optional seal wash for high-salt mobile phases keeps maintenance to a minimum and reduces operating costs.

G7114A Agilent 1260 Infinity II Variable Wavelength Detector
Designed for highest optical performance, GLP compliance and easy maintenance with:
• higher data rate up to 120 Hz
• deuterium lamp for highest intensity and lowest detection limit over awavelength range of 190 to 600 nm
• optional flow-cell cartridges (standard 10 mm, 14 μL; high pressure 10 mm,14 μL; micro 3 mm, 2 μL; semi-micro 6 mm, 5 μL), Bio, and Prep Cells are available and can be used depending on the application needs
• Dual wavelength mode
• easy front access to lamp and flow cell for fast replacement,
• electronic identification of flow cell and lamp with RFID (Radio Frequency Identification) tag for unambiguous identification,
• lamp information: part number, serial number, production date, ignitions, burn time
• cell information: part number, serial number, production date, nominal path length, volume, maximum pressure
• built-in electronic temperature control (ETC) for improved baseline stability
• built-in holmium oxide filter for fast wavelength accuracy verification.

G7116A Agilent 1260 Infinity II Multicolumn Thermostat
• Superior usability with flexible flap positions: open door to 90° (desk function), or to 180°, or even removal of the front door for maximum accessibility.
• Efficient, fast, and most convenient column exchange through InfinityLab Quick Connect Fittings.
• Reproducible precolumn solvent thermostatting with next generation InfinityLab Quick Connect Heat Exchanger easily installed for each column. A bio-inert heat exchanger is available for biological and extreme pH applications.
• Maximum application flexibility through Peltier cooling and heating with two independent temperature zones from 10 °C below ambient temperature up to 85 °C.
• Optional, traceable column identification to track history of up to four columns for documentation of column type, major column parameters, and number of injections as standard for GLP.
• Optional valve drive to host user-exchangeable InfinityLab Quick Change Valve Heads of different formats, which are also available in bio-inert materials.

G7121A Agilent 1260 Infinity II Fluorescence Detector
• Lowest limits of detection with a Raman S/N > 3000 (using dark signal noise reference). Simplified optical design for optimized baseline stability.
• Up to 100 % resolution gain in fast LC using a 74 Hz data acquisition rate.
• Long-life xenon lamp for highest sensitivity. The long-life (> 4000 hours) flash lamp, lamp reference system and efficient light collection ensure constant lamp energy for maximum excitation of fluorophores.
• Easy front access enables fast inspection or exchange of the flow cell.
• Automatic recognition of all flow cell cartridges provides documentation of instrument parameters and helps to comply with GLP.
• Extensive diagnostics, error detection and display with Instant Pilot controller and Agilent Lab Advisor software.
设备使用相关说明
提供HPLC级别甲醇,乙腈,异丙醇和超纯水,特殊试剂需自备。
HPLC-grade methanol, acetonitrile, isopropanol and ultrapure water are provided, special reagents need to be self-prepared.

凡涉及安全风险的样品(如生物危害、危险化学品、人体来源样品、纳米材料、放射性材料等)须通过生物科学中央实验室准入审查;未通过审查不得入内。相关合规义务及由此产生的责任由本人及所属机构自行承担。
Any samples involving safety risks (such as biohazards, hazardous chemicals, human-derived samples, nanomaterials, or radioactive materials) must pass the BioCRF's access review; samples that do not pass the review may not be brought into the BioCRF. Compliance obligations and any resulting liabilities shall be borne by me and my affiliated institution.
备注
提供HPLC级别甲醇,乙腈,异丙醇和超纯水,特殊试剂需自备。
HPLC-grade methanol, acetonitrile, isopropanol and ultrapure water are provided, special reagents need to be self-prepared.
预约资源
附件下载
公告
同类仪器