纳米粒度及Zeta电位分析仪Nano Particle Size and High Sensitive Zeta Potential Analyzer
-
47/人使用者
-
655/次机时次数
-
889/小时总时长
-
0/次送样次数
-
100/人收藏者
-
收费标准
机时0元/小时 -
设备型号
NanoBrook Omni -
当前状态
-
管理员
陈俊毅 Jerry Junyi CHEN 020-88338592 -
放置地点
E4-126/133走廊-127
- 仪器信息
- 预约资源
- 附件下载
- 公告
- 同类仪器
名称
纳米粒度及Zeta电位分析仪Nano Particle Size and High Sensitive Zeta Potential Analyzer
资产编号
BC-GE-0024
型号
NanoBrook Omni
规格
产地
美国
厂家
Brookhaven
所属品牌
Brookhaven Instruments
出产日期
购买日期
所属单位
生物科学中央实验室 Biosciences Central Research Facility (GZ)
使用性质
科研
所属分类
BioCRF-通用仪器设备平台 General Equipment Core
资产负责人
--
联系电话
020-88338592
联系邮箱
jerryjyc@hkust-gz.edu.cn
放置地点
E4-126/133走廊-127
- 主要规格&技术指标
- 主要功能及特色
主要规格&技术指标
1. General Specifications:
*Laser: Standard Laser 40 mW red diode laser, nominal 640 nm wavelength;
*Condensation Control: Purge facility using dry air or nitrogen;
*Temperature Control: -5 °C to 110 °C, active control. No external circulator required;
2. Specifications for Particle Sizing at 90°:
*Sample Type: Most nanoparticle, and colloidal-sized materials, in any non-absorbing liquid;
*Size Range: 1 to 3 mL disposable plastic, 50 µL disposable, 40 µL quartz flow cell, 10 µl quartz minimum;
*Concentration Range: 2 ppm to 50 mg/mL;
*Signal Processing: Dynamic Light Scattering (DLS);
*Correlator: Brookhaven’s TurboCorr, multitau, research grade with 510 hardware channels, covering the equivalent of 1010 linearly-spaced channels, 100% efficiency, real-time operation over the entire delay-time range;
*Precision: ± 1% typically;
*Scattering Angle: 90°;
*Data Presentation: Average & width, lognormal fit, and multimodal size distribution standard;
*Test Standards: Conforms to ISO13321 and ISO22412;
3. Specifications for Particle Sizing at 173°:
*Sample Type: Most globular proteins, nanoparticles, and small polymers in any non-absorbing liquid;
*Size Range: <0.3 nm to 10 µm diameter;
*Sample Cells: 50 µl disposable, 1 to 3 mL disposable plastic, glass, or quartz cells;
*Concentration Range: 0.1 ppm to 50 mg/mL;
4. Specifications for Zeta Potential:
*Sample Type: Most nanoparticle, polymer and colloidal-sized materials, suspended in any non-absorbing liquid, with relative permittivity > 20 and viscosity 1.5 and viscosity < 30 cP PALS;
*Size Range Suitable for Zeta Measurement: 1 nm to 100 µm;
*Mobility Range: 10(9) to 10(7) m2 /V•s ELS; 10(11) to 10(7) m2 /V•s PALS;
*Zeta Potential Range: -500 to 500 mV;
*Sample Cells: 180 µL, 600 µL, 1250 µL;
*Maximum Sample Conductivity: 220 mS/cm;
*Signal Processing: Sizing: Dynamic Light Scattering (DLS), Zeta Potential: Electrophoretic & true Phase Analysis Light Scattering (ELS & PALS);
*Precision: ± 3%;
*Scattering Angle: 15°
*Laser: Standard Laser 40 mW red diode laser, nominal 640 nm wavelength;
*Condensation Control: Purge facility using dry air or nitrogen;
*Temperature Control: -5 °C to 110 °C, active control. No external circulator required;
2. Specifications for Particle Sizing at 90°:
*Sample Type: Most nanoparticle, and colloidal-sized materials, in any non-absorbing liquid;
*Size Range: 1 to 3 mL disposable plastic, 50 µL disposable, 40 µL quartz flow cell, 10 µl quartz minimum;
*Concentration Range: 2 ppm to 50 mg/mL;
*Signal Processing: Dynamic Light Scattering (DLS);
*Correlator: Brookhaven’s TurboCorr, multitau, research grade with 510 hardware channels, covering the equivalent of 1010 linearly-spaced channels, 100% efficiency, real-time operation over the entire delay-time range;
*Precision: ± 1% typically;
*Scattering Angle: 90°;
*Data Presentation: Average & width, lognormal fit, and multimodal size distribution standard;
*Test Standards: Conforms to ISO13321 and ISO22412;
3. Specifications for Particle Sizing at 173°:
*Sample Type: Most globular proteins, nanoparticles, and small polymers in any non-absorbing liquid;
*Size Range: <0.3 nm to 10 µm diameter;
*Sample Cells: 50 µl disposable, 1 to 3 mL disposable plastic, glass, or quartz cells;
*Concentration Range: 0.1 ppm to 50 mg/mL;
4. Specifications for Zeta Potential:
*Sample Type: Most nanoparticle, polymer and colloidal-sized materials, suspended in any non-absorbing liquid, with relative permittivity > 20 and viscosity 1.5 and viscosity < 30 cP PALS;
*Size Range Suitable for Zeta Measurement: 1 nm to 100 µm;
*Mobility Range: 10(9) to 10(7) m2 /V•s ELS; 10(11) to 10(7) m2 /V•s PALS;
*Zeta Potential Range: -500 to 500 mV;
*Sample Cells: 180 µL, 600 µL, 1250 µL;
*Maximum Sample Conductivity: 220 mS/cm;
*Signal Processing: Sizing: Dynamic Light Scattering (DLS), Zeta Potential: Electrophoretic & true Phase Analysis Light Scattering (ELS & PALS);
*Precision: ± 3%;
*Scattering Angle: 15°
主要功能及特色
The NanoBrook Series size and zeta potential analyzers incorporate all you need for fast, routine, sub-micron measurements of size and zeta potential. Based on the principles of Dynamic Light Scattering (DLS) for particle sizing and distribution, and based on doppler velocimetry (electrophoretic light scattering or ELS) for zeta potential measurement, most measurements only take a minute or two. The instrument also includes Phase Analysis Light Scattering (PALS) measurements for samples with low mobilities.
With the NanoBrook Omni, measure samples in nearly every possible suspension environment from high salts such as PBS and sea water to organic solvents and aqueous solutions. Particle and protein sizing, with the highest sensitivity, utilize three measurement angles for truly unbiased measurement results.
Typical application:
Proteins, IgG, peptides, RNA/DNA
Liposomes, exosomes, and other biocolloids
Polysaccharides
Nanoparticles
Polymer latexes
Pharmaceutical preparations
Micelles
Oil/Water and Water/Oil emulsions
Paints and pigments
Colloids
Polymers
Pigments, inks and toners
Cosmetic Formulation
Ceramics and refractories
Emulsions (foodstuffs, cosmetics)
Wastewater treatment monitoring
Carbon blacks
With the NanoBrook Omni, measure samples in nearly every possible suspension environment from high salts such as PBS and sea water to organic solvents and aqueous solutions. Particle and protein sizing, with the highest sensitivity, utilize three measurement angles for truly unbiased measurement results.
Typical application:
Proteins, IgG, peptides, RNA/DNA
Liposomes, exosomes, and other biocolloids
Polysaccharides
Nanoparticles
Polymer latexes
Pharmaceutical preparations
Micelles
Oil/Water and Water/Oil emulsions
Paints and pigments
Colloids
Polymers
Pigments, inks and toners
Cosmetic Formulation
Ceramics and refractories
Emulsions (foodstuffs, cosmetics)
Wastewater treatment monitoring
Carbon blacks
预约资源
附件下载
公告
同类仪器